LIFEPAK 500 automated external defibrillator Modes of Operation
5-17
Service Mode Page 2 of 5
To place the LIFEPAK 500 AED in Service Mode:
1. Unplug all components connected to the LIFEPAK 500 AED
Patient Connector.
2. Simultaneously hold down the
UP ARROW, RIGHT ARROW, and SHOCK keys when
you turn on the AED.
3. When prompted to enter a passcode, enter the passcode key sequence—
UP, ANALYZE (or MENU), RIGHT, RIGHT, SHOCK, ANALYZE (or MENU).
4. If a calibration meter is connected when the passcode is entered, the AED
will go into Energy Calibration Mode.
WARNING!
Shock hazard. Use the Service Mode to disable Auto Battery Test Mode
before disassembling the case. With Auto Battery Test enabled, the AED
automatically charges to 50 joules above any capacitor remnant charge
upon power up after you install a new battery.
5. If a calibration meter is not connected when the passcode is entered, press
the
ANALYZE (or MENU) key to advance to the AUTO BATTERY TEST screen.
The current selection, ON, flashes. Press the
UP ARROW key to toggle
between selections (ON/OFF).
AUTO BATTERY TEST
ON/OFF