LIFEPAK 500 automated external defibrillator Modes of Operation
5-24
Clear Fault Codes
To clear fault codes from the fault code log:
1. Unplug all components connected to the LIFEPAK 500 AED patient
connector.
2. Place the AED in Service Mode.
3. Press the
ANALYZE (or MENU) key until the fault codes are displayed.
4. Press the
UP ARROW key to scroll through the fault codes until END is
displayed.
5. Record all of the displayed fault codes and the date/time each occurred.
6. Press the
ANALYZE (or MENU) key until the CLEAR FAULT CODES? screen is
displayed.
7. Select YES by pressing the
UP ARROW key, then press the ANALYZE (or MENU)
key.
8. Turn off the AED to exit Service Mode.
9. Turn on the AED and verify that the service indicator is off. If the service
indicator remains on, check for new fault codes and make sure the Auto
Battery Test option is turned on. The service indicator will light if there are
faults in the fault code log or when the Auto Battery Test option is turned off.
CLEAR FAULT CODES?
YES/NO