LIFEPAK 500 automated external defibrillator Troubleshooting
8-4
Fault Code Table Page1of20
Code Fault Description
Corrective
Action
Code
5100 boot_fault_minor 10
5101 cannot_determine_date_for_autotest 9
5102 insufficient_shutdown_time 10, 2
5240 eeprom_fault_minor 6, 2
5241 eat_bad_setup_features_crc 6, 2
5242 eat_bad_modem_data_crc 6, 2
5243 eat_bad_calconstants_crc 6, 2
5244 eat_bad_dutycycles_crc 6, 2
5245 eat_bad_shockcountdata_crc 6, 2
5246 eat_bad_operatingtime_crc 6, 2
5247 eat_bad_manufacturingdata_crc 6, 2
5248 eat_bad_initdata_crc 6, 2
5249 eat_bad_aed_mode_crc 6, 2
524a eat_bad_aed_state_crc 6, 2
524b eat_bad_cmonstatedata_crc 6, 2
524c eat_bad_episodeshockcount_crc 6, 2
524d eat_bad_selectedenergy_crc 6, 2
524e eat_bad_seriesshockcount_crc 6, 2
Corrective Action Code Key
Code Description
1 Try to reload the software
2 Replace the Main PCB
3 Replace the Connector
4 Replace the Energy Storage Capacitor
5 Replace the Inductor
6 Recalibrate/Restore Setup, Complete PIP,
Return to service if fault codes do not recur
7 Check the Wires
8 Check the Battery
9 Replace the Time-of-Day Clock Battery
10 Clear Faults, Complete PIP, Return to service
if fault codes do not recur
11 Replace the Biphasic PCB