LIFEPAK 500 automated external defibrillator Replacement Procedures
11-20
Preparing Top Cover for Replacement
Application Case replacement requires preparation of the top cover with labels and other
hardware prior to installation of major components.
To prepare the top cover:
1. Clean the adhesive mating surfaces for the keypanel and the warning/danger
labels with isopropyl alcohol. Allow the surfaces to dry completely.
2. Install the LCD shield. Do not touch the visible surfaces of the shield.
3. Install the keypanel label into the top cover indexing recess. Press in the
label from the center toward the edges. Make sure the keypanel label
adheres firmly to the top cover.
4. Install the warning/danger label into the indexing recess at the front of the
top cover.
5. Install the Physio icon label into the indexing recess in the lower right-hand
corner of the top cover.