LIFEPAK 500 automated external defibrillator Troubleshooting
8-25
Troubleshooting Chart Page 2 of 8
Area Observation Suggested Corrective Action
PowerOn Nopoweron
Use a fully functional battery.
Check installation of ON/OFF key switch cap.
Check battery connection to Main PCB
J12.
Check battery pins.
Check battery polarity.
Reload operating system (Physio-Control
only).
Check or replace the Main PCB.
Improper power on sequence Verify instructions and retry test.
Check or replace Time-of-Day Clock
Battery.
Check or replace Main PCB.
Keyboard Test Cannot enter Service Mode Verify instructions and retry test.
Check installation of key switch caps.
Check keypad overlay.
Verify that the top cover is properly
installed.
Check or replace Main PCB.