LIFEPAK 500 automated external defibrillator Replacement Procedures
11-23
Reassembling Case Page 1 of 2
To reassemble the case:
1. Verify the switch caps are installed and properly aligned and all wiring is
clear of the enclosure and cannot be pinched.
2. Hold the front of the upper enclosure at a 45-degree angle to the front of the
lower enclosure, then lower the upper enclosure.
3. Align the three plastic interlocks in the handle of the upper enclosure with the
corresponding interlocks in the handle of the lower enclosure and carefully
lower the top cover into position.
4. Install 12 new enclosure screws.
Complete the Test and Calibration Procedures (TCP), followed by the
Performance Inspection Procedure (PIP), if you replaced any of the following
components:
■ Case
■ Main PCB
■ Energy Storage Capacitor
■ Wave Shaping Inductor
(LIFEPAK 500 Monophasic
AED only)
■ Battery Cable Harness
■ Biphasic PCB
(LIFEPAK 500 Biphasic AED only)
■ Inductive Resistor
(LIFEPAK 500 Biphasic AED only)
■ Patient Connector Harness