LIFEPAK 500 automated external defibrillator Replacement Procedures
11-10
Recording User-Defined Setups Page 1 of 6
Operating system versions through 4.40 require that you restore user-defined
and service setups in SETUP MODE whenever you reload or upgrade operating
system software. With Monophasic version 4.41 and later and all Biphasic
operating system versions, the loader program creates a temporary image of
setups and restores them when you upgrade software. In some cases, this
temporary recording and restoring of setups may be available in a service
situation; however, it is imperative that service personnel manually check
settings after performing service procedures.
You should record existing user-defined and service setups whenever you intend
to replace the Main PCB assembly. Failure to do so will result in loss of user-
defined and service setups when the AED is returned to use.
A User Defined Setups Log Sheet is provided for your convenience. Print the
Defined Setups Log Sheet before recording user-defined setups. Use the
Defined Setups Log Sheet to record the setups recovered below.
To record user-defined setups:
1. Make sure the LIFEPAK 500 AED power is off.
2. Hold down the
ANALYZE (or MENU), UP ARROW, and RIGHT ARROW keys. Then,
press the
ON/OFF key. Do not release the ANALYZE (or MENU), UP ARROW, and
RIGHT ARROW keys until the SETUP MODE message appears.