LIFEPAK 500 automated external defibrillator Replacement Procedures
11-33
Reseating Main PCB Page 1 of 2
Application The following procedure describes moving the Main PCB back into position
when reassembling the LIFEPAK 500 AED. If the Main PCB is completely
detached, perform the appropriate procedure:
Replacing Main PCB—Monophasic AED
Installing Main PCB—Monophasic AED
Replacing Biphasic PCB—Biphasic AED
Installing Main PCB—Biphasic AED
To reseat the Main PCB:
1. Check mounting grommets on the Main PCB. If any are damaged or missing,
replace them from the Main PCB Installation Kit.
2. Check the five switch caps and LCD board on the top side of the Main PCB.
If necessary, make adjustments or replace damaged parts from the Main
PCB Installation Kit. Check that the battery wire harness is in place.
3. Reconnect P9 of the speaker cable to J9 on the right underside of the
Main PCB.
4. (LIFEPAK 500 Biphasic AED only): reconnect P11 of the flex cable to J11.