LIFEPAK 500 automated external defibrillator Troubleshooting
8-3
Corrective Action Codes
Corrective action codes are listed in the Fault Code Table in the order in which
they should be applied to resolve a specific fault. For example, fault code 5721
suggests two corrective action codes: CAC 7 and CAC 2. The service person
would first inspect the AED wiring associated with the fault, per CAC 7. If wiring
is found to be intact, the Main PCB should be replaced, per CAC 2.
Corrective
Action
Code
Description
1 Try to reload the software
2 Replace the Main PCB
3 Replace the Connector
4 Replace the Energy Storage Capacitor
5 Replace the Inductor
6 Recalibrate/Restore Setup, Complete PIP, Return to service if fault codes do not recur
7 Check the Wires
8 Check the Battery
9 Replace the Time-of-Day Clock Battery
10 Clear Faults, Complete PIP, Return to service if fault codes do not recur
11 Replace the Biphasic PCB