LIFEPAK 500 automated external defibrillator Replacement Procedures
11-5
Warnings and Cautions
The following general warnings and cautions apply to all actions you may
perform during maintenance of the LIFEPAK 500 AED. Other, more specific,
warnings and cautions appear throughout this electronic service manual.
WARNING!
Shock hazard. Servicing of this AED must be performed by properly trained
individuals. This AED may retain potentially lethal charges accessible
inside the AED at any time–even when off. Follow the procedures carefully
for discharging the Energy Storage Capacitor, and disabling Auto Battery
Test.
Shock hazard. The Energy Storage Capacitor carries high voltage up to
several hours after charging. Remove the battery and discharge the
capacitor manually before proceeding to access internal components.
Possible shock and AED damage. Unless reassembled properly, it is
possible to pinch and damage wires during reassembly. To avoid pinching
wires, carefully follow reassembly instructions.
CAUTION!
Possible component damage. The Main PCB assembly contains static
sensitive devices (SSDs). To avoid damage, observe the special handling
practices described in the section titled Static Sensitive Device Handling.