LIFEPAK 500 automated external defibrillator Replacement Procedures
11-4
Introduction
The Replacement Procedures are a set of detailed instructions for disassembly,
handling, and reassembly of replaceable LIFEPAK 500 AED assemblies. The
procedures apply to both the LIFEPAK 500 Monophasic AED and LIFEPAK 500
Biphasic AED, with differences between the two models specifically noted.
Whenever the LIFEPAK 500 AED case must be opened for service, be sure to
observe precautions for preventing shock and protecting electrostatic-discharge-
sensitive components. Review Warnings and Cautions before proceeding.
Inspect the exterior of the LIFEPAK 500 AED for improper mechanical function
and bent, broken, or corroded connectors or pins. Pick up and turn over the AED
and listen for loose or rattling hardware. Locate any loose or rattling hardware
and tighten or replace it.
When disconnecting cables and wire harnesses, label the cables and
connections so that they match easily during reassembly (e.g., J1, J3, etc.).
Each “P” connection mates to a corresponding “J” connector with the same
number. For example, P3 mates with J3. Refer to the LIFEPAK 500 AED
Interconnection Diagram—Monophasic AED or LIFEPAK 500 AED
Interconnection Diagram—Biphasic AED for additional connection
information.