LIFEPAK 500 automated external defibrillator Battery Maintenance
10-16
AED Cleaning Page 3 of 3
Internal Cleaning
Procedures
Clean the interior of the LIFEPAK 500 AED as described below.
WARNING!
Shock hazard. The Energy Storage Capacitor carries high voltage. Remove
the battery and discharge the capacitor before handling.
1. Brush interior surfaces and parts with a nonmetallic, soft-bristle brush.
2. Remove loosened dirt and dust using a dry, low-pressure compressed
(60 psi) air or vacuum cleaner.
3. Wipe metal surfaces with a soft, nonabrasive cloth that has been dampened
with isopropyl alcohol.