LIFEPAK 500 automated external defibrillator Modes of Operation
5-16
Service Mode Page 1 of 5
Service Mode functions shown on the front panel allow qualified personnel to:
■ Turn Auto Battery Test on or off
■ Review and clear the recorded fault codes
■ Review and clear the shock counters
■ Review and clear the hour meter
■ Set the AED serial number
■ Set the AED dash number (LIFEPAK 500 Monophasic AED, operating
system software version 4.42 and earlier only)
■ Review AED prompts
■ Review AED tones
■ Adjust LCD contrast (LIFEPAK 500 Monophasic AED, operating system
software version 4.42 and earlier only)
WARNING!
Shock hazard. Follow the procedures carefully for discharging the Energy
Storage Capacitor whenever opening the LIFEPAK 500 AED case. With
Auto Battery Test enabled, the AED automatically charges to 50 J above
any capacitor remnant charge when powered up in the Service Mode.