LIFEPAK 500 automated external defibrillator Modes of Operation
5-15
Auto Battery Test Mode Page 3 of 3
To force the LIFEPAK 500 AED to perform an Auto Battery Test:
1. Perform a Test Load Test, as described in the LIFEPAK 500 AED Operating
Instructions, to dissipate any capacitor remnant charge.
2. Turn off the AED and remove the battery.
3. Install the battery that you want to test. Wait at least one minute before
turning the AED back on.
4. Turn on the AED and wait until it displays the BATTERY OK message and
then displays CONNECT ELECTRODES. At this point, the AED will have
performed an Auto Battery Test. The BATTERY OK message may not
appear every time the unit is powered on. As long as the battery indicator
does not indicate either a low battery or replace battery condition, you may
proceed.
5. Turn off the AED.