LIFEPAK 500 automated external defibrillator Replacement Procedures
11-11
Recording User-Defined Setups Page 2 of 6
3. The AED displays the SETUP MODE screen (shown at the left). The string
nnnnnnnnnnnnnnnnnnn appears at the top of a printed report and can be
used as a verification of some of the setup and service settings.
4. Press the
ANALYZE (or MENU) key to advance to the DEVICE ID screen.
Record the device ID.
5. Press the
ANALYZE (or MENU) key to advance to the MODEM PHONE
NUMBER screen. Record the modem phone number.
6. Press the
ANALYZE (or MENU) key to advance to the MODEM SELECTION
screen. Record the modem selection.
7. Press the
ANALYZE (or MENU) key to advance to the MODEM INIT STRING
screen. Record the modem init string, if applicable.
Note: The MODEM INIT STRING only displays for a modem selection of 0.
8. Press the
ANALYZE (or MENU) key to advance to the ENERGY SEQUENCE
screen. Record the energy sequence value.
9. Press the
ANALYZE (or MENU) key to advance to the ENERGY PROTOCOL
screen. Record the energy protocol setting.
10. Press the
ANALYZE (or MENU) key to advance to the DISPLAY ENERGY
screen. Record the display energy setting.
SETUP MODE
nnnnnnnnnnnnnnnnnnn