LIFEPAK 500 automated external defibrillator Replacement Procedures
11-12
Recording User-Defined Setups Page 3 of 6
11. (3-button models only:) Press the ANALYZE key to advance to the AUTO
ANALYZE screen. Record the auto analyze setting.
12. Press the ANALYZE (or MENU) key to advance to the CPR TIME screen. Record
the CPR time value. (With monophasic version 4.33 and later and all
biphasic operating system software, record CPR TIME 1 and CPR TIME 2.)
13. (Monophasic operating system version 5.5 and biphasic version 3.8 or later:)
Press the
ANALYZE (or MENU) key to advance to the PULSE PROMPT screen.
Record the setting.
14. (Monophasic operating system version 4.2 and later/all biphasic versions:)
Press the
ANALYZE (or MENU) key to advance to the CPSS DURING CPR
screen. Record the CPSS During CPR setting.
Note: The CPSS During CPR option should always be set to OFF unless
authorized by the Medical Director of the controlling institution.
15. Press the ANALYZE (or MENU) key to advance to the MOTION DETECTION
screen. Record the motion detection setting.
Note: The motion detection option should always be set to ON unless
authorized by the Medical Director of the controlling institution.