EasyManua.ls Logo

Physio Control LIFEPAK 500 - Page 123

Physio Control LIFEPAK 500
259 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Previous Page Table of Contents Section Contents
Back
Index Next Page
LIFEPAK 500 automated external defibrillator Replacement Procedures
11-13
Recording User-Defined Setups Page 4 of 6
16. (Monophasic operating system version 4.33 and later/all biphasic versions:)
Press the
ANALYZE (or MENU) key to advance to the ASYSTOLE DETECTOR
screen. Record the asystole time setting.
Note: The asystole detector option should always be set to OFF unless
authorized by the Medical Director of the controlling institution.
17. (Units enabled for recording only.) Press the ANALYZE (or MENU) key to
advance to the AUDIO RECORDING screen. Record the value.
18. (Monophasic operating system version 4.2 and later/all biphasic versions:)
Press the
ANALYZE (or MENU) key to advance to the PAPER SIZE screen.
Record the paper size value.
19. (Monophasic operating system version 4.5 and later/all biphasic versions:)
Press the
ANALYZE (or MENU) key to advance to the INCIDENT ID screen.
20. (Biphasic version 3.8 or later:) Press the
ANALYZE (or MENU) key to advance to
the AUDIO ALERT screen. Record the setting.
21. (Monophasic operating system version 4.2 and later/all biphasic versions:)
Press the
ANALYZE (or MENU) key to advance to the TRANSFER SETUP
screen.
22. Turn off the LIFEPAK 500 AED.

Table of Contents

Related product manuals