LIFEPAK 500 automated external defibrillator Replacement Procedures
11-14
Recording User-Defined Setups Page 5 of 6
23. Unplug all components connected to the LIFEPAK 500 AED patient connector.
24. Place the AED in Service Mode.
25. Notice that the AED displays the AUTO BATTERY TEST screen. This option
should always be set to ON when the AED is in active service.
26. Press the
ANALYZE (or MENU) key to advance to the FAULT CODE screen.
Record each fault code and the time it occurred.
27. Press the
ANALYZE (or MENU) key to advance to the SHOCK COUNTER, 360 J
screen. Record both shock tallies for the 360 J energy setting (XXXXX
represents the number of shocks since the last counter reset, YYYYY
represents the number of shocks since the unit was built, and ZZZ
represents the energy setting). Press the
UP ARROW key to view and record
the shock tallies for the remaining energy settings.
28. Press the
ANALYZE (or MENU) key to advance to the HOURS screen. Record
both hours values (XXXXX represents the number of hours since the last
hour meter reset and YYYYY represents the number of hours since the unit
was built).
SHOCK COUNTER
XXXXX/YYYYY ZZZJ
HOURS
XXXXX/YYYYY