LIFEPAK 500 automated external defibrillator Replacement Procedures
11-22
Removing Top Cover Page 2 of 2
After removing the top cover, perform the procedure Discharging Energy
Storage Capacitor, or if you have already discharged the capacitor, jump to any
of the following procedures:
■ Reassembling Case
■ Replacing Time-of-Day Clock Battery
■ Replacing a Main PCB Switch Cap
■ Replacing Main PCB—Monophasic AED or Replacing Main PCB—
Biphasic AED
■ Replacing Battery Plate Wire Harness
■ Removing Energy Storage Capacitor—Monophasic AED or Removing
Energy Storage Capacitor—Biphasic AED
■ Replacing Biphasic PCB—Biphasic AED
■ Replacing Speaker Assembly
■ Replacing Wave Shaping Inductor—Monophasic AED or Replacing
Inductive Resistor—Biphasic AED
■ Replacing Patient Connector Harness—Monophasic AED or Replacing
Patient Connector Harness—Biphasic AED
■ Replacing Lower Enclosure Perimeter Seal
■ Replacing Lower Case Corner and Patient Connector Door
■ Replacing Readiness Display—Biphasic AED