LIFEPAK 500 automated external defibrillator Replacement Procedures
11-24
Reassembling Case Page 2 of 2
Complete the Restoring User-Defined Setups procedure if you replaced the
Main PCB.
Complete the Performance Inspection Procedure (PIP) if you replaced any of the
following components:
Any of the components that require you to complete
the Test and Calibration Procedures (TCP)
Time-of-Day Clock Battery
Readiness Display
Any of the switch caps
Speaker Assembly
Lower Enclosure Perimeter Seal