LIFEPAK 500 automated external defibrillator Troubleshooting
8-24
Troubleshooting Chart Page 1 of 8
Area Observation Suggested Corrective Action
Physical Inspection Loose or broken hardware
Locate and tighten or replace loose items.
Locate and replace broken components.
Damaged label or overlay
Replace label or overlay.
Broken or cracked case component
Replace case component.
Damaged or contaminated patient
connector
Clean or replace patient connector.