LIFEPAK 500 automated external defibrillator Troubleshooting
8-24
Troubleshooting Chart Page 1 of 8
Area Observation Suggested Corrective Action
Physical Inspection Loose or broken hardware
■ Locate and tighten or replace loose items.
■ Locate and replace broken components.
Damaged label or overlay
■ Replace label or overlay.
Broken or cracked case component
■ Replace case component.
Damaged or contaminated patient
connector
■ Clean or replace patient connector.