LIFEPAK 500 automated external defibrillator Replacement Procedures
11-19
Preparing Lower Case for Replacement
Application Case replacement requires preparation of the lower case with labels and other
hardware prior to installation of major components.
To prepare the lower case:
1. Clean the adhesive mating surfaces for the serial number label, precautions
label, mounting feet, and speaker felt with isopropyl alcohol. Allow the
surfaces to dry completely.
2. Install the new serial number label, precautions label, and four new feet as
shown below. Install two pieces of new speaker felt.
3. Proceed to Replacing Lower Case Corner and Patient Connector Door.
Serial Number Label
Precautions Label