EasyManua.ls Logo

Physio Control LIFEPAK 500 - Page 129

Physio Control LIFEPAK 500
259 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Previous Page Table of Contents Section Contents
Back
Index Next Page
LIFEPAK 500 automated external defibrillator Replacement Procedures
11-19
Preparing Lower Case for Replacement
Application Case replacement requires preparation of the lower case with labels and other
hardware prior to installation of major components.
To prepare the lower case:
1. Clean the adhesive mating surfaces for the serial number label, precautions
label, mounting feet, and speaker felt with isopropyl alcohol. Allow the
surfaces to dry completely.
2. Install the new serial number label, precautions label, and four new feet as
shown below. Install two pieces of new speaker felt.
3. Proceed to Replacing Lower Case Corner and Patient Connector Door.
Serial Number Label
Precautions Label

Table of Contents

Related product manuals