LIFEPAK 500 automated external defibrillator Modes of Operation
5-18
Service Mode Page 3 of 5
6. Press the ANALYZE (or MENU) key to advance to the FAULT CODES screen.
Press the
UP ARROW key to scroll through the recorded fault codes in
sequence. END signifies there are no more fault codes to review. If there are
no fault codes present, the FAULT CODES screen displays NONE. Click
Clear Fault Codes for more information.
7. The CLEAR FAULT CODES screen only displays when fault codes are
present. The current selection, NO, flashes. Press the
UP ARROW key to toggle
between selections (YES/NO). To complete this procedure, press the
ANALYZE
(or MENU)key.
8. Press the
ANALYZE (or MENU) key to advance to the SHOCK COUNTER
screen. This screen displays the number of shocks the AED has delivered
since the last time the shock counter was reset and the number of shocks
the AED has delivered since it was built. The shock counter screen displays
the number of recorded shocks for each of four energy settings (360J, 300J,
200J, and 5J). XXXXX represents the number of shocks since the last
counter reset, YYYYY represents the number of shocks since the unit was
built, and ZZZ represents the energy setting. The 5J energy setting is for test
load tests only. Press the
UP ARROW key to scroll through the shocks recorded
for each of the energy settings.
FAULT CODES
CODE DDMMMYY HH:MM
CLEAR FAULT CODES?
YES/NO
CLEAR SHOCK COUNTS?
YES/NO
SHOCK COUNTER
XXXXX/YYYYY ZZZJ