LIFEPAK 500 automated external defibrillator Modes of Operation
5-19
Service Mode Page 4 of 5
9. Press the ANALYZE (or MENU) key to advance to the CLEAR SHOCK
COUNTS? screen. The current selection, NO, flashes. Press the
UP ARROW
key to toggle between selections (YES/NO). To complete this procedure,
press the
ANALYZE (or MENU)key.
10. Press the
ANALYZE (or MENU) key to advance to the HOURS screen. This
screen displays the number of hours the AED has been in service since it
was built and since the last time the hours counter was reset (XXXXX
represents the number of hours since the last hour meter reset and YYYYY
represents the number of hours since the unit was built).
11. Press the
ANALYZE (or MENU) key to advance to the CLEAR HOUR METER?
screen. The current selection, NO, flashes. Press the
UP ARROW key to toggle
between selections (YES/NO). To complete this procedure, press the
ANALYZE
(or MENU)key.
12. Press the
ANALYZE (or MENU) key to advance to the SERIAL NUMBER screen.
This screen displays a flashing underscore followed by a field of up to 17
additional characters. Press the
UP ARROW key to scroll through the entire
character set. Press the
RIGHT ARROW key to scroll through the 18 positions in
the serial number character field.
13. (LIFEPAK 500 Monophasic AED, operating system 4.42 or earlier only):
Press the
ANALYZE (or MENU) key to advance to the DASH NUMBER screen.
Record the dash number.
CLEAR SHOCK COUNTS?
___
HOURS
XXXXX/YYYYY
CLEAR HOUR METER?
YES/NO
SERIAL NUMBER
_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _