LIFEPAK 500 automated external defibrillator Modes of Operation
5-20
Service Mode Page 5 of 5
14. Press the ANALYZE (or MENU) key to advance to the PROMPT REVIEW
screen. Press the
UP ARROW key to scroll through all of the AED screen
messages and voice prompts. For a complete listing of screen messages
and voice prompts, refer to the LIFEPAK 500 AED Operating Instructions.
15. Press the
ANALYZE (or MENU) key to advance to the TONE REVIEW screen.
Press the
UP ARROW key to scroll through all of the AED tones.
CAUTION!
Possible device malfunction. (Monophasic software version 4.42 and
earlier): Incorrect adjustment of LCD contrast could affect the LCD’s
readability at temperature extremes of 0° and 50°C (32° and 122°F). Contact
Physio-Control for further information.
16. (Monophasic software version 4.42 and earlier): Press the
ANALYZE (or
MENU) key to advance to the ADJUST LCD CONTRAST screen. Press the UP
ARROW
key to scroll through available values for LCD contrast (the default
value is 60%).
17. Turn power off at any time to exit Service Mode.
Note: New settings are saved only if the
ANALYZE (or MENU) key is pressed
after changing an entry. If you press the
ANALYZE (or MENU) key after you have
exited Service Mode, the display advances to the AUTO BATTERY TEST
screen again.