LIFEPAK 500 automated external defibrillator Product Description
3-11
Functional Description Page 2 of 2
Patient Connector A port used for patient connection, data
communications, and unit testing.
Keypad Function keys (ON/OFF, ANALYZE [in 2-button units, blank “menu” key],
SHOCK, UP ARROW, and RIGHT ARROW) that enable operator input. The UP ARROW and
RIGHT ARROW keys are used for Time-of-Day Clock setup, Service Mode, and
Data Transfer Mode.
ANALYZE is used manually to initiate patient evaluation
(3-button models only). The
SHOCK key is used to deliver defibrillation energy.
Combinations of these keys are used to place the AED in various operating
modes.
LCD—Atwo-line, 20-character, dot matrix LCD used to display operator
prompts, system status messages, and AED configuration information.
Readiness Display A 3-field LCD used to present Service and Battery
indicators’ information for edge-on viewing when the AED is secured in a
stationary mount (biphasic models only).
Speaker Used by the AED to annunciate voice prompts and tones.
Microphone Part of the Main PCB, the microphone picks up scene audio
data, such as operator remarks, and AED voice prompts or tones for
inclusion in the patient record (scene-audio recording models only).