LIFEPAK 500 automated external defibrillator Troubleshooting
8-22
Fault Code Table Page 19 of 20
Code Fault Description
Corrective
Action
Code
9b0b contin_stack_too_small_tshutdown 10
9b0c contin_stack_too_small_tdsp_comm 10
9b0d contin_stack_too_small_tuart_comm_recv 10
9b0e contin_stack_too_small_tdeterminecurrentkeystate 10
9b0f contin_stack_too_small_tdistrib 10
9b10 contin_stack_too_small_tuart_comm_xmit 10
9b11 contin_stack_too_small_ttrans 10
9b12 contin_stack_too_small_tapp 10
9b13 contin_stack_too_small_tdefib 10
9b14 contin_stack_too_small_tuserinterface_task 10
9b15 contin_stack_too_small_tcontrol_aed_task 10
9b16 contin_stack_too_small_tsas 10
9b17 contin_stack_too_small_teeprom 10
9b18 contin_stack_too_small_tfaults 10
9b19 contin_stack_too_small_tpres 10
9b1a contin_stack_too_small_tarch 10
9b1b contin_stack_too_small_tcontinmonitor 10
9d00 exception_fault_major 10
9e00 bus_fault_major 10
Corrective Action Code Key
Code Description
1 Try to reload the software
2 Replace the Main PCB
3 Replace the Connector
4 Replace the Energy Storage Capacitor
5 Replace the Inductor
6 Recalibrate/Restore Setup, Complete PIP,
Return to service if fault codes do not recur
7 Check the Wires
8 Check the Battery
9 Replace the Time-of-Day Clock Battery
10 Clear Faults, Complete PIP, Return to service
if fault codes do not recur
11 Replace the Biphasic PCB