LIFEPAK 500 automated external defibrillator Battery Maintenance
10-3
Battery Maintenance Page 2 of 12
Rechargeable
Battery Pak
(continued)
Recharge until the battery charger charge LED is green. This indicates that
the battery charger has completed the fast-charge cycle. Under charging can
cause battery damage.
Recharge only at temperatures between +15° and +35°C (+59° and +95°F).
Do not expose battery paks to temperatures greater than +50°C (+122°F).
Do not allow electrical connection between the battery contacts.
WARNING!
Possible loss of power during patient care. Using an improperly
maintained battery to power the LIFEPAK 500 AED may cause premature
power loss. Use only the LIFEPAK 500 AED battery charger to charge the
LIFEPAK 500 rechargeable battery pak.
CAUTIONS!
Possible battery damage. Recharge the battery until the battery charger
charge LED is green. Under charging can cause battery damage.
Possible battery damage. Charging batteries outside the temperature
range of +15° to +35°C (+59° to +95°F) may cause improper charging and
shorten battery life.
Possible battery damage. Electrical connection between battery contacts
can blow an internal fuse and permanently disable the battery.