LIFEPAK 500 automated external defibrillator Battery Maintenance
10-5
Battery Maintenance Page 4 of 12
Nonrechargeable
Battery Pak
(continued)
All personnel should familiarize themselves with the nonrechargeable battery
information in the LIFEPAK 500 AED Operating Instructions.
To properly maintain nonrechargeable lithium battery paks:
Do not attempt to recharge (lithium battery paks cannot be connected to the
battery charger used to recharge the rechargeable SLA battery paks).
Do not use beyond the expiration date marked on the battery label.
Do not expose to temperatures greater than +50°C (+122°F).
Do not allow electrical connections between the battery contacts.
WARNING!
Possible explosion, fire, or n oxious gas. Attempting to recharge a LIFEPAK
500 nonrechargeable lithium battery pak can cause an explosion or fire or
release noxious gas. Dispose of expired or depleted lithium battery paks
as described in these maintenance instructions.
CAUTION!
Possible battery damage. Electrical connection between battery contacts
can blow an internal fuse and permanently disable the battery.