Remote Control
R&S
®
FPC
229User Manual 1178.4130.02 ─ 07
Example: //Abort calibration
CAL:STAR? S11C
'Connect Open to Port 1'
CAL:ABOR
Usage: Event
Manual operation: See "Calibrate" on page 397
CALibration<n>:CONTinue?
This command resumes calibration after a calibration standard has been connected.
Suffix:
<n>
.
irrelevant
Return values:
<Instructions> String containing the instructions for the next calibration step.
You can find a list possible instructions in the table below.
Example: //Resume calibration
CAL:STAR? S11C
'Connect Open to Port 1'
//Connect Open
CAL:CONT?
'Connect Short to Port 1'
Usage: Query only
Manual operation: See "Calibrate" on page 397
Table 19-6: Return values: Possible instructions for calibration
'Connect Open to Port 1' Connect the open to port 1 before resuming calibration.
'Connect Short to Port 1' Connect the short to port 1 before resuming calibration.
'Connect Load to Port 1' Connect the load to port 1 before resuming calibration.
'Connect Through' Connect a through to ports 1 and 2.
'Disconnect DUT from port 1'
'Calibration Done' Calibration has been finished. No further actions required.
CALibration<n>:MODE?
This command queries the current calibration type.
Suffix:
<n>
.
irrelevant
Return values:
<State> 0
Not calibrated or default calibration.
1
User calibration.
Calibration