EasyManua.ls Logo

R&S FPC Series - Page 277

R&S FPC Series
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FPC
274User Manual 1178.4130.02 ─ 07
If necessary, the corresponding delta marker is activated first.
Suffix:
<n>
.
irrelevant
<m> Marker
Parameters:
<Position> <numeric value>
The unit depends on the measurement and marker input mode.
Hz for measurements in the frequency domain.
s for measurements in the time domain.
No unit for channel input mode (CALCulate<n>:MARKer<m>:
FREQuency:MODE).
Example: //Delta marker positions are relative to marker 1.
CALC:DELT:MOD REL
//Position delta marker 2 10.7 MHz to the right of marker 1.
CALC:DELT2:X 10.7MHz
//Query the absolute and relative position of delta marker 2.
CALC:DELT2:X?
CALC:DELT2:X:REL?
Manual operation: See "Marker <x>" on page 374
See "Marker <x>" on page 384
See "Marker <x>" on page 395
See "Marker <x>" on page 403
CALCulate<n>:DELTamarker<m>:X:RELative <Position>
This command positions a delta marker on a position relative to the reference marker.
If necessary, the corresponding delta marker is activated first.
Suffix:
<n>
.
irrelevant
<m> Marker
Parameters:
<Position> <numeric value>
The unit depends on the measurement and marker input mode.
Hz for measurements in the frequency domain.
s for measurements in the time domain.
No unit for channel input mode (CALCulate<n>:MARKer<m>:
FREQuency:MODE).
Example: //Set the delta marker at a distance of 5 kHz to the reference
position.
CALC:DELT3:X:REL 5 kHz