Remote Control
R&S
®
FPC
351User Manual 1178.4130.02 ─ 07
STATus:QUEStionable:FREQuency[:EVENt]?...................................................................353
STATus:QUEStionable:LIMit[:EVENt]?..............................................................................353
STATus:QUEStionable:POWer[:EVENt]?..........................................................................353
STATus:QUEStionable[:EVENt]?......................................................................................353
STATus:QUEStionable:FREQuency:CONDition?...............................................................353
STATus:QUEStionable:LIMit:CONDition?..........................................................................353
STATus:QUEStionable:POWer:CONDition?......................................................................353
STATus:QUEStionable:CONDition?..................................................................................353
STATus:QUEStionable:FREQuency:ENABle.....................................................................353
STATus:QUEStionable:LIMit:ENABle................................................................................353
STATus:QUEStionable:POWer:ENABle............................................................................353
STATus:QUEStionable:ENABle....................................................................................... 353
STATus:QUEStionable:FREQuency:NTRansition.............................................................. 354
STATus:QUEStionable:LIMit:NTRansition......................................................................... 354
STATus:QUEStionable:POWer:NTRansition......................................................................354
STATus:QUEStionable:NTRansition.................................................................................354
STATus:QUEStionable:FREQuency:PTRansition...............................................................354
STATus:QUEStionable:LIMit:PTRansition......................................................................... 354
STATus:QUEStionable:POWer:PTRansition......................................................................354
STATus:QUEStionable:PTRansition................................................................................. 354
FORMat:SREGister <Format>
This command selects the data format for all queries of the status register.
Parameters:
<Format> ASCii
Returns the data as ASCii bytes in <NR1> format.
BINary
Returns the data in binary format, preceded by '#B'.
HEXadecimal
Returns the data in hexadecial format, preceded by '#H'.
OCTal
Returns the data in octal format, preceded by '#Q'.
*RST: ASCii
Example: //Select data format for status register queries
FORM:SREG BIN
STATus:PRESet
This command resets the edge detectors and ENABle parts of all registers to a defined
value. All PTRansition parts are set to FFFFh, i.e. all transitions from 0 to 1 are detec-
ted. All NTRansition parts are set to 0, i.e. a transition from 1 to 0 in a CONDition bit is
not detected. The ENABle part of the STATus:OPERation and STATus:QUEStionable
registers are set to 0, i.e. all events in these registers are not passed on.
Example:
//Reset edge detectors and ENABle parts of all registers
STAT:PRES