Remote control basics
R&S
®
NRPxxP
134User Manual 1179.5760.02 ─ 02
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
+
Calibrating
0
0
0
Measuring
Triggering
0
0
0
0
Sense summary
0
0
0
0
0
Figure 10-7: Operation status register
Querying the register:
STATus:OPERation:CONDition?
STATus:OPERation[:EVENt]?
Table 10-9: Used operation status bits and their meaning
Bit no. Short description Bit is set if
0 Calibrating Summary of Operation calibrating status register exists.
4 Measuring Summary of Operation measuring status register exists.
5 Triggering Summary of Operation trigger status register exists.
10 Sense summary Summary of Operation sense status register exists.
10.2.5.1 Operation calibrating status register
The CONDition register contains information whether a power sensor is being calibra-
ted. The EVENt register contains information whether a calibration was started or com-
pleted since the last query.