ZVA/ZVB/ZVT SOURce<Ch>:POWer:CORRection...
1145.1084.12 6.275 E-6
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:AVERage:NTOLerance
<tolerance>
Specifies the maximum deviation of the measured power from the target power of the calibration.
<Ch>
Calibrated channel number
<Pt>
Calibrated port number. This parameter is ignored; the tolerance
value is valid for all sources.
<tolerance>
Tolerance value
Range [def. unit] 0.001 dB to 1000 dB [dB]
*RST value 1 dB
SCPI, Command
Types
Device-specific, command or query
Example:
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:DATA '<src_string>', <block> |
<numeric_value>...
Reads or writes source power correction data sets. A power correction data set contains n real values
where:
Each value corresponds to the ratio of the actual power at the reference plane (value provided
by the used source) to the uncalibrated power in dB.
The number n is equal to the number of sweep points.
Increasing (decreasing) the values in the correction data sets increases (decreases) the power at the
reference plane. Writing correction data (the setting command) fails if the number of transferred values
is not equal to the number of sweep points.
<Ch>
Calibrated channel number
<Pt>
Calibrated port number
'<src_string>'
Identifier for the source of the calibrated wave:
• 'A<n>' or 'PORT<n>' denote correction data for the
analyzer port no. <n>.
• 'ESRC<n>' or 'GEN<n>' denote correction data for the
external generator no. <n>.
<block> |
<numeric_value>...
Power correction values in block data format or as a comma-
separated list of real numbers. The query returns a list of
numbers.
Range [def. unit] All numbers are interpreted as dB-values.
*RST value –
SCPI, Command
Types
Device-specific, command or query