SOURce<Ch>:POWer:CORRection... R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.274 E-6
Example:
SOUR:POW3:CORR:COLL BSEN
Perform a source power calibration for port 3 using power meter
no. 2.
SOUR:POW:CORR:PMET:ID 2,
Select power meter no. 2.
SOUR:POW:CORR:ACQ PORT,3
Perform a source power calibration for port 3 using the previously
selected power meter no. 2.
SOURce<Ch>:POWer<Pt>:CORRection:COLLect[:ACQuire]:VERification[:STATe
] <Boolean>
Enables or disables a verification sweep that the analyzer performs after the source power calibration.
The command is valid for all ports and external generators
<Ch>
Calibrated channel number
<Pt>
Calibrated port number. This parameter is ignored; the verification is
either enabled or disabled for all sources.
<Boolean>
Enables (ON) or disables (OFF) the verification sweep.
*RST value ON
SCPI,
Command
Types
Device-specific, command or query
Example:
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:AVERage[:COUNt] <readings>
Sets a limit for the number of calibration sweeps in the source power calibration. The command is valid
for all ports and external generators. Equivalent command:
SOURce<Ch>:POWer<Pt>:CORRection:NREadings.
<Ch>
Calibrated channel number
<Pt>
Calibrated port number. This parameter is ignored; the verification is
either enabled or disabled for all sources.
<readings>
Number of readings
Range [def.
unit]
1 to 100 [–]
*RST value 1
SCPI,
Command
Types
Device-specific, command or query
Example:
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire] (for equivalent
command SOURce<Ch>:POWer<Pt>:CORRection:NREadings).