Remote Control Commands
R&S
®
Scope Rider RTH
390User Manual 1326.1578.02 ─ 12
METer<m>:ABORt
Stops the running measurement.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Usage: Event
15.7.5 Results
Reading Values..................................................................................................... 390
METer:MEASure Commands................................................................................391
Statistics................................................................................................................393
15.7.5.1 Reading Values
After configuration and selection of the active measurement, you can read the result
values.
METer<m>:INITiate........................................................................................................ 390
METer<m>:READ?.........................................................................................................390
METer<m>:FETCh?....................................................................................................... 391
METer<m>:INITiate
Resets the hardware and all statistical values, performs a single measurement and
ends in hold mode.
Suffix:
<m>
.
1..4
Usage: Event
METer<m>:READ?
Starts a new measurement, returns the current measurement result and stays in run
mode.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Return values:
<Meter result>
Usage: Query only
Digital Multimeter (R&S RTH1002)