Remote Control Commands
R&S
®
Scope Rider RTH
393User Manual 1326.1578.02 ─ 12
Query parameters:
[<ExpectedValue>] <numeric value> | MIN | MAX | DEF | 'AUTO'
Optional parameter, expected measurement result or range. The
instrument sets the appropriate measurement range.
See: METer:CONFigure:VOLTage:DC.
Return values:
<Result> Numeric value
Example:
:METer:MEASure:VOLTage:DC? 5
<-- 3.4907681509
Usage: Query only
15.7.5.3
Statistics
Before you can get statistical results, configure the measurement and select the mea-
surement to be performed using METer:SENSe:FUNCtion.
METer<m>:CALCulate:AVERage:AVERage?.................................................................... 393
METer<m>:CALCulate:AVERage:MINimum?.................................................................... 393
METer<m>:CALCulate:AVERage:MAXimum?................................................................... 394
METer<m>:CALCulate:AVERage:CLEar...........................................................................394
METer<m>:CALCulate:AVERage:AVERage?
Returns the mean value of the measurement series.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Return values:
<Average> Range: -100E+24 to 100E+24
Increment: 1E-12
*RST: 0
Default unit: V
Usage: Query only
METer<m>:CALCulate:AVERage:MINimum?
Returns the minimum value of the measurement series.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.