EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - Page 392

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
Scope Rider RTH
392User Manual 1326.1578.02 ─ 12
METer:MEASure:CAPacitance? [<ExpectedValue>]
Configures the capacitance measurement and returns the result.
Query parameters:
[<ExpectedValue>] <numeric value> | MIN | MAX | DEF | 'AUTO'
Optional parameter, expected measurement result or range. The
instrument sets the appropriate measurement range.
See: METer:CONFigure:CAPacitance.
Return values:
<Result> Numeric value
Usage: Query only
METer:MEASure:CURRent:AC?
[<ExpectedValue>]
METer:MEASure:CURRent:DC? [<ExpectedValue>]
Configures the current measurement and returns the result.
Query parameters:
[<ExpectedValue>] <numeric value> | MIN | MAX | DEF | 'AUTO'
Optional parameter, expected measurement result or range. The
instrument sets the appropriate measurement range.
See: METer:CONFigure:CURRent:DC.
Return values:
<Result> Numeric value
Example:
:METer:MEASure:CURRent:DC? 15
<-- 13.4907681509
Usage: Query only
METer:MEASure:RESistance? [<ExpectedValue>]
Configures the resistance measurement and returns the result.
Query parameters:
[<ExpectedValue>] <numeric value> | MIN | MAX | DEF | 'AUTO'
Optional parameter, expected measurement result or range. The
instrument sets the appropriate measurement range.
See: METer:CONFigure:RESistance.
Return values:
<Result> Numeric value
Usage: Query only
METer:MEASure:VOLTage:AC? [<ExpectedValue>]
METer:MEASure:VOLTage:DC? [<ExpectedValue>]
Configures the voltage measurement and returns the result.