Command Description R&S SMB
1407.0806.32 6.6 E-2
CALibration Subsystem
The CALibration system contains the commands for adjustment. Adjustment is triggered by the query
commands. The response "0" indicates error-free adjustment, and the response "1" means that an error
occurred during adjustment.
Command Parameters Default unit Remark
CALibration:ALL[:MEASure]? Query only
CALibration:FMOFfset[:MEASure]? Query only
CALibration:FREQuency[:MEASure]? Query only
CALibration:LEVel:STATe ON | OFF
CALibration:LEVel[:MEASure]? Query only
CALibration:ALL[:MEASure]?
The command starts all internal adjustments for which no external measuring equipment is needed.
The command is a query command and therefore has no *RST value.
Example:
CAL:ALL?
'starts the adjustment of all functions for the entire instrument.
Response: 0
'adjustment has been performed successfully.
*RST value Resolution Options SCPI
- - Compliant
CALibration:FMOFfset[:MEASure]?
The command starts all adjustment for the FM/PhiM modulator.
The command is a query command and therefore has no *RST value.
Example:
CAL:FMOF?
'starts the adjustments for the FM/PhiM modulator.
Response: 0
'the adjustments have been performed successfully.
*RST value Resolution Options SCPI
- - Device-specific
CALibration:FREQuency[:MEASure]?
The command starts all adjustments which affect the frequency.
The command is a query command and therefore has no *RST value.
Example:
CAL:FREQ?
'starts the adjustments for maximum frequency accuracy.
Response: 0
'the adjustments have been performed successfully.
*RST value Resolution Options SCPI
- - Device-specific