5. Measurements with the Barracuda System
Introduction
Barracuda & QABrowser Reference Manual
2012-10/4.3A
110
Very Low Dose Rate Measurements with EMM-BiasW5.2
To get access to the wide range of the
EMM-BiasW, update mode Free run or
Timed needs to be selected, as shown to the
left. Both these modes measure without any
trig levels or background compensation. See
topic Update Modes for more
information.
Then "Very high" sensitivity must be chosen,
as shown in the next figure. This is done
under "Settings - XXX", where XXX is the
detector you have connected to your
EMM-BiasW module.