2. Description of the Barracuda
Hardware and Specifications
2012-10/4.3A
19
Barracuda & QABrowser Reference Manual
Typically the exposure time has to be at least 5 ms to get a kVp value but it depends
on the waveform. On modern X-ray generators (high-frequency with fast rise and fall
times) the peak tube voltage can normally be measured with exposure time as short as
1 ms. Dose and time values will be given for even shorter exposure times.
The multi-purpose detector is connected to the multi-purpose module with a 2 m cable
(an optional 6 m cable is also available), however all specifications given are for the
2 m cable.
2.6.2.2 Specifications, MPD
The inaccuracy is here defined as the root of the square sum of systematic errors,
which has not been eliminated, and random errors (dispersion around a mean value).
The calculation of the inaccuracy is based on 15 different measurements and with a
confidence level of 95 %. Of the total inaccuracy, random error is 20 % and general
inaccuracy is 80 %.
Note: Irradiation time is often called exposure time in daily use.
Reference conditions