5. Measurements with the Barracuda System
Radiography
Barracuda & QABrowser Reference Manual
2012-10/4.3A
116
5.3.2 Dose Measurements with R100 or Ion Chamber
If the R100B is selected all types of electrometer modules can be used. For an
ionization chamber only one of the EMM-Bias electrometer modules can be used,
since it supports bias to an ion chamber.
1. Place the R100 or Ion Chamber in the field and connect
the cable to the EMM input.
2. Set up the Barracuda and the handheld computer
according to the description in Setting Up the
Barracuda .
3. Follow the same steps as for the measurement with
MPD, but select Dose as parameter, the EMM module
instead of the MPM module, and an ionization chamber
or R100B from the detector list.
The EMM LED is flashing to assist you to confirm that
your detector has been connected to the module of
your choice.