5. Measurements with the Barracuda System
Cine/Pulsed Radiography
2012-10/4.3A
121
Barracuda & QABrowser Reference Manual
5.4.1 kVp, Time, Dose, and Dose Rate
Use the same procedure as for the normal radiography measurement but select the
All... parameter.
If only Tube voltage is selected:
5.4.2 Pulse Measurements with R100 or Ion Chamber
Use the same procedure as for the normal radiography measurement using ionization
chamber. Note that some ion chambers cannot be used to measure pulses since they
have too slow response. In that case use a manual pulse rate setting, as described
under Settings to get a dose per pulse reading.
62