EasyManua.ls Logo

RTI Barracuda - 5.4.3 HVL, Quick-HVL, and Total Filtration; 5.5.1 Image Intensifier Input Dose Rate

RTI Barracuda
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Measurements with the Barracuda System
Fluoroscopy and Pulsed Fluoroscopy
2012-10/4.3A
123
Barracuda & QABrowser Reference Manual
Normally the R100 is used together with the
electrometer module to be able to measure
the lowest possible dose rate levels down to
0.1 µGy/s. Another reason to use the R100 is
that the detector is much smaller than the
MPD making it easier to position in front of the
image intensifier without affecting the
measuring field for the mA feedback loop.
If the image intensifier manually can control the mA and kV, then you can use the MPD
for measurements down to about 0.7 µGy/s. For pulsed fluoroscopy even lower levels
can be measured.
As a secondary parameter the total dose is accumulated. After you have turned off the
fluoroscopy unit, this value is used to calculate the average dose rate as total dose
divided by the exposure time.
Note that for very low dose rate values the exposure time cannot be measured
accurately and the last dose rate value cannot be stored automatically in the display.
Then tap Hold to "freeze" the current value in the display. The waveform is also
acquired when you tap Hold. Waveform is also automatically acquired when the
selected delay time expires.
Select "I.I. input dose rate" as measuring parameter to be able to measure lowest
possible dose rate and tap Hold to "freeze" current value in the display. The total
accumulated dose is shoved after you have switched off the fluoroscopy unit.
5.5.1 Image Intensifier Input Dose Rate
Use patient equivalent phantom to measure the image intensifier input dose rate
according to manufacturer's specification:
1. Connect the R100 to the EMM.
2. Set up the Barracuda and the handheld computer according to the description in
Setting Up the Barracuda .
3. Select type of measurement.
16

Table of Contents