EasyManuals Logo

RTI Barracuda User Manual

RTI Barracuda
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
2. Description of the Barracuda
Maintenance
2012-10/4.3A
49
Barracuda & QABrowser Reference Manual
Next, the available detectors
are shown to the left.
When clicking a detector, the
available calibrations will
show up to the right (In this
case the MPD is
highlighted). The TV and TF
columns show an × if there
are calibrations for Tube
Voltage and/or Total
Filtration. The factor column
shows the calibration factor
(for dose in this case).
Here is another example
(R100). This type of detector
only contains a calibration
factor for dose.
2.8.4 Transferring Detector Calibrations
For the Barracuda, the calibration factors for the external detectors are stored inside
the cabinet.
If you have access to more than one cabinet you will need to transfer the detector
calibrations to the other cabinet(s) to be able to use all available detectors with all the
systems.
To transfer a detector:
1. Connect the cabinet, containing the calibrations, to a PC.
2. Start the RTI Detector Manager program (see Managing Detector Calibrations ).
3. Choose Main | File | Export.
48

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RTI Barracuda and is the answer not in the manual?

RTI Barracuda Specifications

General IconGeneral
BrandRTI
ModelBarracuda
CategoryMedical Equipment
LanguageEnglish