3. Description of the QABrowser
QABrowser Applications
2012-10/4.3A
73
Barracuda & QABrowser Reference Manual
3.4.1 The Accuracy Application (single-parameter)
1. Go to the real-time display that displays
only kVp. Tap Appl or press the
corresponding button to open the
application list.
It is recommended to make one exposure
first to analyse the waveform and estimate
the total filtration.
2. The Select application screen lists the
available applications for selected Type of
measurement and Selected parameter.
For kVp, only Accuracy and
Reproducibility are available.
Choose Accuracy by tapping it with the
pen. You can also use the scroll button to
highlight Accuracy and then press the
button that corresponds to Select (the
right-most button).
3. The accuracy application is shown on the
screen. The set values (for kVp) are
stored in a Set-value list. You can modify
the list or individual values.
To modify an individual set value tap with
the pen on it. In this case tap 60.