Barracuda & QABrowser Reference Manual
The MAS-2B probe is mostly used for
mAs measurements since it is not
sensitive enough to measure tube
current on fluoroscopy. The lowest tube
current that can be measured with
MAS-2B is 10 mA. The figure to the left
shows the MAS-2B probe without the
cable. Read the MAS-2B User's Manual
for a detailed description and instructions
on how to set up and connect it.
The parameter mAs is available for most type of measurements but mAs is normally
measured only for X-ray exposures.
To measure tube charge (mAs) with the MAS-2B
1. Connect the MAS-2B probe to the X-ray generator as described in the MAS-2B
User's Manual.
2. Connect the MAS-2B adapter cable to the EMM.
3. Set up the Barracuda and the handheld computer according to the description in
Setting Up the Barracuda .
4. In the QABrowser select type of
measurement.
5. Next select parameter mAs from the
menu.
The Select Detector screen is now
displayed. Select the MAS-2B probe, and
tap Select.
6. The real-time display for mAs is now displayed. Set the generator. Tap Reset. Make
an exposure.
16