EasyManuals Logo

RTI Barracuda User Manual

RTI Barracuda
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #40 background imageLoading...
Page #40 background image
2. Description of the Barracuda
Hardware and Specifications
Barracuda & QABrowser Reference Manual
2012-10/4.3A
36
Module type
Range
Inaccuracy
Typ. noise
current.
Note 1: For Tube current the specified Inaccuracy is valid for Irradiation times > 100 ms. See
Note for R100B above.
- MAS-2B, Non-invasive mAs probe (sensitivity 1 nC/mAs)
Module type
Range
Inaccuracy
Typ. noise
EMM-1Ch/2Ch/
Bias/BiasB/BiasW
Tube charge
Tube current
Pulse tube current
0.1 mAs
10 4000 mA
Lower limit 50 mA,
otherwise same as for
tube current.
±5 %
±5 % or ±2 mA (±3 %
at 250 mA)
±1 mA
Note 1: For Tube current the specified Inaccuracy is valid for Irradiation times > 20 ms. See Note
for R100B above.
- MAS-3, Non-invasive mAs probe (sensitivity -1 nC/mAs)
Module type
Range
Inaccuracy
Typ. noise
EMM-1Ch/2Ch/
Bias/BiasB
Tube charge
Tube current
Pulse tube current
0.001 mAs
0.1 – 2000 mA
Lower limit 1 mA,
otherwise same as for
tube current.
±1 %
±1 % or ±0.02 mA
±5 µA
EMM-BiasW
N/A
N/A
N/A
Note 1: For Tube current the specified Inaccuracy is valid for Irradiation times > 20 ms. See Note
for R100B above.
- L100, Light detector (typical sensitivity -670 pA/nit or -200 pA/lx)
Module type
Range
Inaccuracy
Typ. noise
EMM-1Ch/2Ch/
Bias/BiasB
Luminance
Illuminance
0.03 – 72000 cd/m²
0.01 – 24000 lx
±5 % or ±0.006 cd/m²
±5 % or ±0.002 lx
±3 mcd/m²
±1 mlx
EMM-BiasW
N/A
N/A
N/A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RTI Barracuda and is the answer not in the manual?

RTI Barracuda Specifications

General IconGeneral
BrandRTI
ModelBarracuda
CategoryMedical Equipment
LanguageEnglish