1. Troubleshooting
8-13
Error
number
Message
Probable cause
Action
084
Retort heating
time-out.
Effect of environmental temperature.
Let the instrument stand for a while at an
operable temperature and then turn on the
power.
Heater or wiring to the heater was
disconnected.
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty heater driver board.
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty CPU board.
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty SSR.
Contact the Sakura Finetek Technical Support
representative or local representative.
085
Rotary valve heating
time-out.
Effect of environmental temperature.
Let the instrument stand for a while at an
operable temperature and then turn on the
power.
Heater or wiring to the heater was
disconnected.
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty heater driver board.
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty CPU board.
Contact the Sakura Finetek Technical Support
representative or local representative.
086
Oven temperature
exceeded 80C.
Hot molten paraffin, etc., was introduced
to the oven.
Do not put hot molten paraffin.
Faulty CPU board.
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty heater driver board.
Contact the Sakura Finetek Technical Support
representative or local representative.
087
Rotary valve
temperature
exceeded 80C.
Hot liquid was introduced into the
instrument.
Do not put hot liquid.
Faulty CPU board.
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty heater driver board.
Contact the Sakura Finetek Technical Support
representative or local representative.
088
Retort temperature
exceeded 85C.
Introduction of hot reagent and warm
water.
Do not put hot liquid.
Power restoration under special
condition.
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty CPU board.
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty SSR.
Contact the Sakura Finetek Technical Support
representative or local representative.
090
Exhaust fan alarm.
Faulty fume fan or disconnected relay
cable.
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty CPU board.
Contact the Sakura Finetek Technical Support
representative or local representative.
091
Cooling fan alarm.
Faulty cooling fan or disconnected relay
cable
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty CPU board.
Contact the Sakura Finetek Technical Support
representative or local representative.
101
CF card not found.
CF is removed or half removed.
Faulty CF
Contact the Sakura Finetek Technical Support
representative or local representative.
Faulty CPU board.
Contact the Sakura Finetek Technical Support
representative or local representative.
110
Low battery.
Consumed battery.
Contact the Sakura Finetek Technical Support
representative or local representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Histo-Tek VP1 and is the answer not in the manual?

Sakura Histo-Tek VP1 Specifications

General IconGeneral
BrandSakura
ModelHisto-Tek VP1
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals