1. Troubleshooting
8-14
Error
number
Message
Probable cause
Action
120 ~ 145
File system error.
No available space in the CF card.
Contact the Sakura Finetek Technical Support
representative or local representative.
Damaged CF card file or file format
error.
Contact the Sakura Finetek Technical Support
representative or local representative.
A specified CF card is not used.
Contact the Sakura Finetek Technical Support
representative or local representative.
150
WAV file read failure.
The CF file system is damaged and
the specified file cannot be found.
Contact the Sakura Finetek Technical Support
representative or local representative.
151
Sound file formatting
error.
The CF file system is damaged and
the file format is different from the one
specified.
Contact the Sakura Finetek Technical Support
representative or local representative.
160 ~ 199
File write error.
File read error.
The specified file is not found in the
CF card.
Contact the Sakura Finetek Technical Support
representative or local representative.
File in the CF card or USB flash drive
is damaged, file format error, or
write-protected file.
Contact the Sakura Finetek Technical Support
representative or local representative.
A specified CF card is not used.
Contact the Sakura Finetek Technical Support
representative or local representative.
200
USB flash drive not
found.
Removed in the middle. Or, the USB
flash drive failed.
Do not remove the USB flash drive while it is being
accessed.
If the USB flash drive is faulty, replace it with a new
one.
210 ~ 213
An error occurred while
software update is in
progress.
There is no software to be updated in
the USB flash drive.
Contact the Sakura Finetek Technical Support
representative or local representative.
File in the CF card is damaged, file
format error, or write-protected file.
Contact the Sakura Finetek Technical Support
representative or local representative.
A specified CF card is not used.
Contact the Sakura Finetek Technical Support
representative or local representative.
250
A power outage has
occurred. Processing
has been aborted due to
the possibility of the
specimens drying.
Power outage continued for 8 minutes
or longer.
Use an uninterruptible power supply (UPS).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Histo-Tek VP1 and is the answer not in the manual?

Sakura Histo-Tek VP1 Specifications

General IconGeneral
BrandSakura
ModelHisto-Tek VP1
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals